Student and Faculty Publications

Publication Date

8-1-2023

Journal

Journal of Interventional Cardiac Electrophysiology

Abstract

Background: Rotating and laser sheaths are both routinely used in transvenous lead extraction (TLE) which can lead to catastrophic complications including death. The efficacy and risk of each approach are uncertain. To perform a meta-analysis to compare success and mortality rates associated with rotating and laser sheaths.

Methods: We searched electronic academic databases for case series of consecutive patients and randomized controlled trials published 1998-2017 describing the use of rotating and laser sheaths for TLE. Among 48 studies identified, rotating sheaths included 1,094 patients with 1,955 leads in 14 studies, and laser sheaths included 7,775 patients with 12,339 leads in 34 studies. Patients receiving rotating sheaths were older (63 versus 60 years old) and were more often male (74% versus 72%); CRT-P/Ds were more commonly extracted using rotating sheaths (12% versus 7%), whereas ICDs were less common (37% versus 42%), p > 0.05 for all. Infection as an indication for lead extraction was higher in the rotating sheath group (59.8% versus 52.9%, p = 0.002). The mean time from initial lead implantation was 7.2 years for rotating sheaths and 6.3 years for laser sheaths (p > 0.05).

Results: Success rates for complete removal of transvenous leads were 95.1% in rotating sheaths and 93.4% in laser sheaths (p < 0.05). There was one death among 1,094 patients (0.09%) in rotating sheaths and 66 deaths among 7,775 patients (0.85%) in laser sheaths, translating to a 9.3-fold higher risk of death with laser sheaths (95% CI 1.3 to 66.9, p = 0.01).

Conclusions: Laser sheaths were associated with lower complete lead removal rate and a 9.3-fold higher risk of death.

Keywords

Humans, Male, Middle Aged, Device Removal, Defibrillators, Implantable, Lasers, Retrospective Studies, Equipment Failure, Pacemaker, Artificial, Treatment Outcome, Transvenous lead extraction, Laser sheaths, Rotating sheaths, CIEDs, Mortality

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.