Abstract
In this work, we integrated a commercially-available fully-spectroscopic pixelated cadmium telluride (CdTe) detector system as a two-dimensional (2D) array detector into our existing benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system. After integrating this detector, known as High-Energy X-ray Imaging Technology (HEXITEC), we performed quantitative imaging of gold nanoparticle (GNP) distribution in a small animal-sized phantom using our benchtop XFCT system. Owing to the upgraded detector component within our benchtop XFCT system, we were able to conduct this phantom imaging in an unprecedented manner by volumetric XFCT scans followed by XFCT image reconstruction in 3D. The current results showed that adoption of HEXITEC, in conjunction with a custom-made parallel-hole collimator, drastically reduced the XFCT scan time/dose. Compared with the previous work performed with our original benchtop XFCT system adopting a single crystal CdTe detector, the currently observed reduction was up to a factor of 5, while achieving comparable GNP detection limit under similar experimental conditions. Overall, we demonstrated, for the first time to the best our knowledge, the feasibility of benchtop XFCT imaging of small animal-sized objects containing biologically relevant GNP concentrations (on the order of 0.1 mg Au/cm
Keywords
X-ray fluorescence, X-ray fluorescence computed tomography, CdTe detector, HEXITEC, gold nanoparticle, quantitative imaging
DOI
10.1109/access.2022.3210590
PMID
36274671
PMCID
PMC9583747
PubMedCentral® Posted Date
10-20-2022
PubMedCentral® Full Text Version
Post-print
Published Open-Access
yes
Included in
Bioinformatics Commons, Biomedical Informatics Commons, Genetic Phenomena Commons, Medical Genetics Commons, Oncology Commons