Background and purpose: A novel cobalt-60 compensator-based intensity-modulated radiation therapy (IMRT) system was developed for a resource-limited environment but lacked an efficient dose verification algorithm. The aim of this study was to develop a deep-learning-based dose verification algorithm for accurate and rapid dose predictions.

Materials and methods: A deep-learning network was employed to predict the doses from static fields related to beam commissioning. Inputs were a cube-shaped phantom, a beam binary mask, and an intersecting volume of the phantom and beam binary mask, while output was a 3-dimensional (3D) dose. The same network was extended to predict patient-specific doses for head and neck cancers using two different approaches. A field-based method predicted doses for each field and combined all calculated doses into a plan, while the plan-based method combined all nine fluences into a plan to predict doses. Inputs included patient computed tomography (CT) scans, binary beam masks, and fluence maps truncated to the patient's CT in 3D.

Results: For static fields, predictions agreed well with ground truths with average deviations of less than 0.5% for percent depth doses and profiles. Even though the field-based method showed excellent prediction performance for each field, the plan-based method showed better agreement between clinical and predicted dose distributions. The distributed dose deviations for all planned target volumes and organs at risk were within 1.3 Gy. The calculation speed for each case was within two seconds.

Conclusions: A deep-learning-based dose verification tool can accurately and rapidly predict doses for a novel cobalt-60 compensator-based IMRT system.

Keywords

Deep-learning, Dose prediction, Fluence map, Cobalt-60 compensator-based IMRT

DOI

10.1016/j.phro.2023.100440

PMID

37342210

PMCID

PMC10277917

PubMedCentral® Posted Date

4-21-2023

PubMedCentral® Full Text Version

Post-print

Published Open-Access

yes

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.