Faculty, Staff and Student Publications
Publication Date
7-1-2024
Journal
The Use of Computers in Radiation Therapy
Abstract
The Dice Similarity Coefficient (DSC) is the current de facto standard to determine agreement between a reference segmentation and one generated by manual / auto-contouring approaches. This metric is useful for non-spatially important images; however, radiation therapy requires consideration of nearby Organs-at-Risk (OARs) and their radiosensitivity which are currently unaccounted for with the traditional DSC. In this work, we introduce the OAR-DSC which accounts for nearby OARs and their radiosensitivity when computing the DSC. We illustrate the importance of this through cases where two proposed contours have similar DSC, but lower OAR-DSC when one contour expands closer to the surrounding OARs. This work is important because the OAR-DSC may be used by deep learning auto-contouring algorithms in a radiation therapy specific loss function, thereby progressing on the current disregard for the importance of these differences on the final radiation dose plan generation, delivery, and risks of patient toxicity.
PMID
39512542
PMCID
PMC11542924
PubMedCentral® Posted Date
11-7-2024
PubMedCentral® Full Text Version
Author MSS
Published Open-Access
yes
Included in
Bioinformatics Commons, Biomedical Informatics Commons, Genetic Phenomena Commons, Medical Genetics Commons, Oncology Commons