Faculty, Staff and Student Publications
Publication Date
1-6-2022
Journal
Environmental Monitoring and Assessment
Abstract
Previous validation studies found a good linear correlation between the low-cost particulate matter sensors (LCPMS) and other research grade particulate matter (PM) monitors. This study aimed to determine if different particle size bins of PM would affect the linear relationship and agreement between the Dylos DC1700 (LCPMS) particle count measurements (converted to PM2.5 mass concentrations) and the Grimm 11R (research grade instrument) mass concentration measurements. Three size groups of PM2.5 (mass median aerodynamic diameters (MMAD): < 1 µm, 1-2 µm, and > 2 µm) were generated inside a laboratory chamber, controlled for temperature and relative humidity, by dispersing sodium chloride crystals through a nebulizer. A linear regression comparing 1-min average PM2.5 particle counts from the Dylos DC1700 (Dylos) to the Grimm 11R (Grimm) mass concentrations was estimated by particle size group. The slope for the linear regression was found to increase as MMAD increased (< 1 µm, 0.75 (R2 = 0.95); 1-2 µm, 0.90 (R2 = 0.93); and > 2 µm, 1.03 (R2 = 0.94). The linear slopes were used to convert Dylos counts to mass concentration, and the agreement between converted Dylos mass and Grimm mass was estimated. The absolute relative error between converted Dylos mass and the Grimm mass was smaller in the < 1 µm group (16%) and 1-2 µm group (16%) compared to the > 2 µm group (32%). Therefore, the bias between converted Dylos mass and Grimm mass varied by size group. Future studies examining particle size bins over a wider range of coarse particles (> 2.5 µm) would provide useful information for accurately converting LCPMS counts to mass concentration.
Keywords
Aerosols, Air Pollutants, Environmental Monitoring, Laboratories, Particle Size, Particulate Matter, Low-cost Sensors, Particulate Matter (PM), PM sensor calibration, Particle size
DOI
10.1007/s10661-021-09715-6
PMID
34989887
PMCID
PMC8842881
PubMedCentral® Posted Date
1-6-2023
PubMedCentral® Full Text Version
Author MSS
Published Open-Access
yes