Publication Date
4-1-2021
Journal
Gastrointestinal Endoscopy
DOI
10.1016/j.gie.2020.07.007
PMID
32682812
PMCID
PMC7855348 D
PubMedCentral® Posted Date
4-1-2022
PubMedCentral® Full Text Version
Author MSS
Published Open-Access
yes
Keywords
Epithelial Cells, Esophagoscopy, Esophagus, Humans, Neoplasms, Sensitivity and Specificity, Software, artificial intelligence, computer-assisted diagnosis, high-resolution microendoscopy, esophageal squamous cell neoplasia, esophageal cancer
Abstract
BACKGROUND AND AIMS: High-resolution microendoscopy (HRME) is an optical biopsy technology that provides subcellular imaging of esophageal mucosa but requires expert interpretation of these histopathology-like images. We compared endoscopists with an automated software algorithm for detection of esophageal squamous cell neoplasia (ESCN) and evaluated the endoscopists' accuracy with and without input from the software algorithm.
METHODS: Thirteen endoscopists (6 experts, 7 novices) were trained and tested on 218 post-hoc HRME images from 130 consecutive patients undergoing ESCN screening/surveillance. The automated software algorithm interpreted all images as neoplastic (high-grade dysplasia, ESCN) or non-neoplastic. All endoscopists provided their interpretation (neoplastic or non-neoplastic) and confidence level (high or low) without and with knowledge of the software overlay highlighting abnormal nuclei and software interpretation. The criterion standard was histopathology consensus diagnosis by 2 pathologists.
RESULTS: The endoscopists had a higher mean sensitivity (84.3%, standard deviation [SD] 8.0% vs 76.3%, P = .004), lower specificity (75.0%, SD 5.2% vs 85.3%, P < .001) but no significant difference in accuracy (81.1%, SD 5.2% vs 79.4%, P = .26) of ESCN detection compared with the automated software algorithm. With knowledge of the software algorithm, the specificity of the endoscopists increased significantly (75.0% to 80.1%, P = .002) but not the sensitivity (84.3% to 84.8%, P = .75) or accuracy (81.1% to 83.1%, P = .13). The increase in specificity was among novices (P = .008) but not experts (P = .11).
CONCLUSIONS: The software algorithm had lower sensitivity but higher specificity for ESCN detection than endoscopists. Using computer-assisted diagnosis, the endoscopists maintained high sensitivity while increasing their specificity and accuracy compared with their initial diagnosis. Automated HRME interpretation would facilitate widespread usage in resource-poor areas where this portable, low-cost technology is needed.