Publication Date

10-1-2023

Journal

Journal of Clinical Monitoring and Computing

DOI

10.1007/s10877-023-00985-8

PMID

36934402

PMCID

PMC10024795

PubMedCentral® Posted Date

3-19-2023

PubMedCentral® Full Text Version

Post-print

Published Open-Access

yes

Keywords

Humans, Brain Injuries, Arterial Pressure, Glasgow Coma Scale, Cerebrovascular Circulation, Intracranial Pressure, Cerebral blood flow, Autoregulation, Intracranial pressure, Rheoencephalography, Noninvasive

Abstract

Cerebral blood flow (CBF) autoregulation (AR) can be monitored using invasive modalities, such as intracranial pressure (ICP) and arterial blood pressure (ABP) to calculate the CBF AR index (PRx). Monitoring PRx can reduce the extent of secondary brain damage in patients. Rheoencephalography (REG) is an FDA-approved non-invasive method to measure CBF. REGx, a CBF AR index, is calculated from REG and arm bioimpedance pulse waves. Our goal was to test REG for neuromonitoring. 28 measurement sessions were performed on 13 neurocritical care patients. REG/arm bioimpedance waveforms were recorded on a laptop using a bioimpedance amplifier and custom-built software. The same program was used for offline data processing. Case #1: The patient's mean REGx increased from - 0.08 on the first day to 0.44 on the second day, indicating worsening intracranial compliance (ICC) (P < 0.0001, CI 0.46-0.58). Glasgow Coma Scale (GCS) was 5 on both days. Case #2: REGx decreased from 0.32 on the first recording to 0.07 on the last (P = 0.0003, CI - 0.38 to - 0.12). GCS was 7 and 14, respectively. Case #3: Within a 36-minute recording, REGx decreased from 0.56 to - 0.37 (P < 0.0001, 95%, CI - 1.10 to - 0.76). Central venous pressure changed from 14 to 9 mmHg. REG pulse wave morphology changed from poor ICC to good ICC morphology. Bioimpedance recording made it possible to quantify the active/passive status of CBF AR, indicate the worsening of ICC, and present it in real time. REGx can be a suitable, non-invasive alternative to PRx for use in head-injured patients.

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.