    Abstract

    The present study aimed to evaluate the biomechanical behavior of PEEK abutments with different heights on single titanium implants. To investigate the implant surface, different tests (scanning electron microscopy, energy-dispersive X-ray, and X-ray diffraction) were adopted. Herein, 20 implants received the 4.5 × 4.0 mm PEEK short abutment (SA) and 20 received the 4.5 × 5.5 mm PEEK long abutment (LA). The abutments were installed using dual-cure resin cement. To determine the fatigue test, two specimens from each group were submitted to the single load fracture test. For this, the samples were submitted to a compressive load of (0.5 mm/min; 30°) in a universal testing machine. For the fatigue test, the samples received 2,000,000 cycles (2 Hz; 30°). The number of cycles and the load test was analyzed by the reliability software SPSS statistics using Kaplan-Meier and Mantel-Cox tests (log-rank) (p < 0.05). The maximum load showed no statistically significant differences (p = 0.189) for the SA group (64.1 kgf) and the LA group (56.5 kgf). The study groups were statistically different regarding the number of cycles (p = 0.022) and fracture strength (p = 0.001). PEEK abutments can be indicated with caution for implant-supported rehabilitation and may be suitable as temporary rehabilitation.

    Keywords

    bioengineering, dental abutments, dental implants, dental materials, polyetheretherketone

    Published Open-Access

    yes

    Included in

    Dentistry Commons

    Share

    COinS
     
     

    To view the content in your browser, please download Adobe Reader or, alternately,
    you may Download the file to your hard drive.

    NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.