Abstract

PURPOSE: Hypoxemia during a failed airway scenario is life threatening. A dual-lumen pharyngeal oxygen delivery device (PODD) was developed to fit inside a traditional oropharyngeal airway for undisrupted supraglottic oxygenation and gas analysis during laryngoscopy and intubation. We hypothesized that the PODD would provide oxygen as effectively as high-flow nasal cannula (HFNC) while using lower oxygen flow rates.

METHODS: We compared oxygen delivery of the PODD to HFNC in a preoxygenated, apneic manikin lung that approximated an adult functional residual capacity. Four arms were studied: HFNC at 20 and 60 liters per minute (LPM) oxygen, PODD at 10 LPM oxygen, and a control arm with no oxygen flow after initial preoxygenation. Five randomized 20-minute trials were performed for each arm (20 trials total). Descriptive statistics and analysis of variance were used with statistical significance of

RESULTS: Mean oxygen concentrations were statistically different and decreased from 97% as follows: 41 ± 0% for the control, 90 ± 1% for HFNC at 20 LPM, 88 ± 2% for HFNC at 60 LPM, and 97 ± 1% (no change) for the PODD at 10 LPM.

CONCLUSION: Oxygen delivery with the PODD maintained oxygen concentration longer than HFNC in this manikin model at lower flow rates than HFNC.

Keywords

Continuous oxygen delivery, high-flow nasal cannula, oropharyngeal airway, pharyngeal oxygen delivery device, supraglottic oxygenation