
Faculty, Staff and Student Publications
Accuracy of IOL Power Calculation Formulas for AcrySof SN60WF versus Tecnis ZCB00 Intraocular Lenses
Publication Date
8-15-2022
Journal
Journal of Ophthalmic and Vision Research (JOVR)
Abstract
PURPOSE: To compare the accuracy of various intraocular lens power formulas for two monofocal hydrophobic foldable lenses, the AcrySof SN60WF and the Tecnis ZCB00.
METHODS: This retrospective study included 409 eyes from 409 patients who underwent uncomplicated cataract surgery (299 eyes with SN60WF and 110 eyes with ZCB00). Biometry was performed for all eyes with an IOLMaster 700. Predicted refraction from five different IOL power formulas (Barrett Universal II, Haigis, Hoffer-Q, Holladay 2, and SRK/T) was compared to postoperative refraction at one to three months for the following axial length strata: short eyes (
RESULTS: In patients with medium eyes, there were no significant differences in the mean absolute error (MAE) and the percentage of eyes within
CONCLUSION: Although there were no significant differences in the formula accuracy between these two lenses in medium eyes for all formulas and in short eyes for most formulas, the accuracy decreased significantly in long eyes for ZCB00 compared to SN60WF. The effect of IOL model on the postoperative outcomes should be further investigated.
Keywords
Formula Accuracy, Intraocular Lens, SN60WF, ZCB00
DOI
10.18502/jovr.v17i3.11571
PMID
36160094
PMCID
PMC9493430
PubMedCentral® Posted Date
August 2022
PubMedCentral® Full Text Version
Post-print
Published Open-Access
yes
Comments
PMID: 36160094