Faculty, Staff and Student Publications

Publication Date

11-1-2022

Journal

The Texas Heart Institute Journal

Abstract

Minimally invasive aortic valve replacement through a right thoracotomy is frequently performed in patients with aortic valve disease. The Cor-Knot Device (LSI Solutions) is an automated fastener that secures valve sutures. This case report is for a patient who developed postcardiotomy shock during a minimally invasive aortic valve surgery. The patient was found to have an aortic root dissection involving 90% of the aortic root circumference, including bilateral coronary ostia. The autopsy revealed that the aortic damage could be explained by a direct aortic intimal tear from the distal tip of the device shaft. The device was most likely not in perfect apposition to the sewing ring because of the restricted angle and space between the ribs.

Keywords

Humans, Aortic Dissection, Aorta, Heart, Thoracotomy, Sutures

Comments

PMID: 36450144

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.