Date of Graduation

8-2011

Document Type

Thesis (MS)

Program Affiliation

Medical Physics

Degree Name

Masters of Science (MS)

Advisor/Committee Chair

Stephen F. Kry, PhD

Committee Member

David Followill, PhD

Committee Member

R. Allen White, PhD

Committee Member

Laurence Court, PhD

Committee Member

Lei Dong, PhD

Committee Member

Adam Garden, MD

Abstract

Purpose: First, to determine an average and maximum displacement of the shoulder relative to isocenter over the course of treatment. Second, to establish the dosimetric effect of shoulder displacements relative to correct isocenter alignment on the dose delivered to the target and the surrounding structures for head and neck cancer patients.

Method and Materials: The frequency of shoulder shifts of various magnitudes relative to isocenter was assessed for 4 patients using image registration software. The location of the center of the right and left humeral head relative to isocenter (usually C2) was found daily from CT on rails scans, and was compared to the location of the humeral heads relative to isocenter on the initial simulation CT. Three Baseline head and neck IMRT and SmartArc plans were generated in Pinnacle based on simulation CTs. The CT datasets (external contour and boney structures) were then modified to represent shifts of the shoulder (relative to isocenter) between 3 mm and 15 mm in the SI, AP, and LR directions. The initial plans were recalculated on the image sets with shifted shoulders.

Results: On average, shoulder variation was 2-5 mm in each direction, although displacements of over 1 cm in the inferior and posterior directions occurred. Shoulder shifts induced perturbations in the dose distribution, although generally only for large shifts. Most substantially, large, superior shifts resulted in coverage loss by the 95% isodose line for targets in the lower neck. Inferior shifts elevated the dose to the brachial plexus by 0.6-4.1 Gy. SmartArc plans showed similar loss of target coverage as IMRT plans.

Conclusions: The position of the shoulder can have an impact on target coverage and critical structure dose. Shoulder position may need to be considered for setup of head and neck patients depending on target location.

Keywords

Head and Neck, shoulder shifts, IMRT, VMAT, SmartArc

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.