Faculty, Staff and Student Publications

Publication Date

5-1-2023

Journal

Journal of Applied Clinical Medical Physics

Abstract

Reject rate analysis is considered an integral part of a diagnostic radiography quality control (QC) program. A rejected image is a patient radiograph that was not presented to a radiologist for diagnosis and that contributes unnecessary radiation dose to the patient. Reject rates that are either too high or too low may suggest systemic department shortcomings in QC mechanisms. Due to the lack of standardization, reject data often cannot be easily compared between radiography systems from different vendors. The purpose of this report is to provide guidance to help standardize data elements that are required for comprehensive reject analysis and to propose data reporting and workflows to enable an effective and comprehensive reject rate monitoring program. Essential data elements, a proposed schema for classifying reject reasons, and workflow implementation options are recommended in this task group report.

Keywords

Humans, Radiography, Quality Control, Reference Standards, Computed radiography, digital radiography, radiography, reject, reject rate analysis, reject rate, repeat