Faculty, Staff and Student Publications

Publication Date

7-1-2024

Journal

Cytometry Part B: Clinical Cytometry

DOI

10.1002/cyto.b.22116

PMID

36824056

Abstract

Detection of measurable residual disease (MRD) in chronic lymphocytic leukemia (CLL) is an important prognostic marker. The most common CLL MRD method in current use is multiparameter flow cytometry, but availability is limited by the need for expert manual analysis. Automated analysis has the potential to expand access to CLL MRD testing. We evaluated the performance of an artificial intelligence (AI)-assisted multiparameter flow cytometry (MFC) workflow for CLL MRD. We randomly selected 113 CLL MRD FCS files and divided them into training and validation sets. The training set (n = 41) was gated by expert manual analysis and used to train the AI model. We then compared the validation set (n = 72) MRD results obtained by the AI-assisted analysis versus those by expert manual analysis using the Pearson correlation coefficient and Bland-Altman plot method. In the validation set, the AI-assisted analysis correctly categorized cases as MRD-negative versus MRD-positive in 96% of cases. When comparing the AI-assisted analysis versus the expert manual analysis, the Pearson r was 0.8650, mean bias was 0.2237 log10 units, and the 95% limit of agreement (LOA) was ±1.0282 log10 units. The AI-assisted analysis performed sub-optimally in atypical immunophenotype CLL and in cases lacking residual normal B cells. When excluding these outlier cases, the mean bias improved to 0.0680 log10 units and the 95% LOA to ±0.2926 log10 units. An automated AI-assisted workflow allows for the quantification of MRD in CLL with typical immunophenotype. Further work is required to improve performance in atypical immunophenotype CLL.

Keywords

Leukemia, Lymphocytic, Chronic, B-Cell, Humans, Neoplasm, Residual, Flow Cytometry, Workflow, Artificial Intelligence, artificial intelligence, automated analysis, chronic lymphocytic leukemia, minimal/measurable residual disease, multiparameter flow cytometry

Published Open-Access

yes

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.