Faculty, Staff and Student Publications

Publication Date

3-1-2023

Journal

Journal of Applied Clinical Medical Physics

Abstract

Purpose: To develop and evaluate an automated whole-brain radiotherapy (WBRT) treatment planning pipeline with a deep learning-based auto-contouring and customizable landmark-based field aperture design.

Methods: The pipeline consisted of the following steps: (1) Auto-contour normal structures on computed tomography scans and digitally reconstructed radiographs using deep learning techniques, (2) locate the landmark structures using the beam's-eye-view, (3) generate field apertures based on eight different landmark rules addressing different clinical purposes and physician preferences. Two parallel approaches for generating field apertures were developed for quality control. The performance of the generated field shapes and dose distributions were compared with the original clinical plans. The clinical acceptability of the plans was assessed by five radiation oncologists from four hospitals.

Results: The performance of the generated field apertures was evaluated by the Hausdorff distance (HD) and mean surface distance (MSD) from 182 patients' field apertures used in the clinic. The average HD and MSD for the generated field apertures were 16 ± 7 and 7 ± 3 mm for the first approach, respectively, and 17 ± 7 and 7 ± 3 mm, respectively, for the second approach. The differences regarding HD and MSD between the first and the second approaches were 1 ± 2 and 1 ± 3 mm, respectively. A clinical review of the field aperture design, conducted using 30 patients, achieved a 100% acceptance rate for both the first and second approaches, and the plan review achieved a 100% acceptance rate for the first approach and a 93% acceptance rate for the second approach. The average acceptance rate for meeting lens dosimetric recommendations was 80% (left lens) and 77% (right lens) for the first approach, and 70% (both left and right lenses) for the second approach, compared with 50% (left lens) and 53% (right lens) for the clinical plans.

Conclusion: This study provided an automated pipeline with two field aperture generation approaches to automatically generate WBRT treatment plans. Both quantitative and qualitative evaluations demonstrated that our novel pipeline was comparable with the original clinical plans.

Keywords

Humans, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiometry, Tomography, X-Ray Computed, Brain, Radiotherapy, Intensity-Modulated, automation, customizable field aperture, deep learning, whole‐brain radiotherapy

DOI

10.1002/acm2.13839

PMID

36412092

PMCID

PMC10018662

PubMedCentral® Posted Date

11-22-2022

PubMedCentral® Full Text Version

Post-print

Published Open-Access

yes

Plum Print visual indicator of research metrics
PlumX Metrics
  • Citations
    • Citation Indexes: 4
  • Captures
    • Readers: 11
  • Mentions
    • News Mentions: 1
see details

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.